7 Operation
7.6 Positioning the patient
I
H
54
Operating Instructions GALILEOS Comfort PLUS (CANADA) GALILEOS
7.6.1
Positioning the patient - with bite block
● Press the R key to move the unit back to the entry position.
● The patient approaches the unit moving backwards.
● If the rotary ring was accidentally displaced while positioning the
patient, the unit can be returned to the entry position by pressing the
"R" key again.
7.6.1.1
Aligning the patient - with standard bite block
● Close the swivel arm (H) until it snaps in place.
● Using the "up" and "down" keys on the control panel, move the X-ray
unit so that the bite block and the patient's front teeth are at the same
height.
The motor movement is accompanied by an acoustic signal.
CAUTION
K
The height adjustment motor starts slowly and then increases its speed.
Press and hold down the height adjustment key until the unit has
reached the desired height.
● The patient then approaches the bite block and holds the handles (I)
firmly.
IMPORTANT
The patient must stand upright with a straight back and relaxed
shoulders.