Figure 10-9 Diode I-V Curve; Figure 10-8 Connections For Diode I-V Tests - Keithley SourceMeter 2400 User Manual

Sourcemeter 2400 series source measure unit
Hide thumbs Also See for SourceMeter 2400:
Table of Contents

Advertisement

®
2400 Series SourceMeter
User's Manual
For the purposes of this test, assume the following basic sweep parameters:
Source Function: current
Sense Function: volts
Source Mode: sweep
Start Current: 1mA
Stop Current: 10mA
Step Current: 1mA
Voltage Compliance: 1V
Source Delay: 100ms
Figure 10-8
shows typical test connections for this test, and
diode curve. Note that the diode anode is connected to INPUT/OUTPUT HI, and the cath-
ode is connected to INPUT/OUTPUT LO. These connections are required to properly for-
ward bias the diode for the purposes of the test. The test connections could also be
reversed by using negative sweep voltage parameters.
Figure 10-8
Connections for diode I-V tests
MEAS
EDIT
V
I
DISPLAY
0
1
TOGGLE
LOCAL
REL
POWER
6
7
DIGITS SPEED
Figure 10-9
Diode I-V curve
V
reverse
-V
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
4 WIRE
SENSE
HI
1100V
5V
PEAK
PEAK
SourceMeter
LO
SOURCE
FCTN
V
I
RANGE
2
3
4
5
EDIT
AUTO
FILTER
LIMIT
TRIG
SWEEP
ON/OFF
RANGE
8
9
+/-
STORE RECALL
CONFIG MENU
EXIT
ENTER
OUTPUT
I
V
V
I
forward
leakage
-I
Sweep Operation
10-23
Figure 10-9
shows a typical
INPUT/
OUTPUT
Diode
Under
1100V
PEAK
Test
250V
!
PEAK
TERMINALS
FRONT/
REAR

Advertisement

Table of Contents
loading

Table of Contents