B) Technical Alarms - Saadat ALBORZ B9 Operator's Manual

Patient monitor

Advertisement

Chapter 5: ECG Monitoring

b) Technical alarms

Alarm
Cause
ECG cable is not
ECG NO
connected to the
CABLE
system
ECG
Mentioned leads are
CHECK
not properly
LA,RA,LL
connected.
ECG
ECG module failure
DEFECT
ECG
RL or other leads are
CHECK RL
not properly
OR ALL
connected.
LL or other leads are
ECG
not properly
CHECK LL
connected when ECG
OR ALL
lead is I for 3wire
lead set
Solution
Connect ECG cable
Make sure that mentioned
electrode is properly connected
Power off and then on the system
.If this message is displayed again
the user should contact local After
Sale Service.
Make sure that all electrodes and
patient cable are properly
connected
Make sure that all electrodes and
patient cable are properly
connected
5-24
User Manual
Explanation
Alarm level 3- the message is
displayed in cyan background. By
pressing ALARM SILENCE,
background becomes gray and
alarm is disabled and ignores this
fault.
Alarm level 2- the message is
displayed in yellow background. By
pressing ALARM SILENCE,
background becomes gray and the
alarm is disabled and ignores this
fault.
Alarm level 2- the message is
displayed in yellow background. By
pressing ALARM SILENCE,
background becomes gray and the
alarm is disabled and ignores this
fault.
Alarm level 2- the message is
displayed in yellow background. By
pressing ALARM SILENCE,
background becomes gray and the
alarm is disabled and ignores this
fault.
Alarm level 2- the message is
displayed in yellow background. By
pressing ALARM SILENCE,
background becomes gray and the
alarm is disabled and ignores this
fault.