Chapter 7: ST Monitoring
NOTE:
ST monitoring is available for adult and pediatric patient and it is not
recommended for neonates.
NOTE:
If there are not at least 5 normal complexes in the last 50 beats of ECG signal, the
ST value will not be displayed.
NOTE:
Applied lead for ST, ARR, Pace and HR is reference lead that is displayed in the
first trace and can be adjusted in ECG menu.
NOTE:
To ensure proper analysis of ST segment deviation, it is recommended to use
Extended filter.
ST monitoring function is "OFF" as a default. You can switch it "ON", when this monitoring is
necessary.
When ST monitoring is enabled, current ST values are trended and can be reviewed on the
TREND window.
NOTE:
Measurement unit of ST segment is "mV".
Measurement range of ST segment is between -2.0 mV to +2.0 mV.
Measurement symbol of ST segment "+" means elevating and "-"means depressing.
ST value in ECG parameters area
7-3
User Manual