Chapter 10: NIBP Monitoring
NIBP UNIT
Pick this item to adjust measurement unit. (Options: mmHg or KPa)
ALARM LEVEL
Selectable between 1 and 2. Level 1 represents the most serious case.
NIBP ALARM
Pick "ON" to enable NIBP alarm functions such as parameters blinking, audio alarm and
light indicator. Pick "OFF" to disable the alarm functions and there will be a "
in the Parameter Area.
NIBP SYS LIMIT
SYS alarm is activated when the systolic pressure exceeds adjusted ALARM HIGH limit
or falls below adjusted ALARM LOW limit.
Adult → Max: 255, Min: 30
Pediatric → Max: 240, Min: 30
Neonate → Max: 135, Min: 30
NIBP DIA LIMIT
DIA alarm is activated when the diastolic pressure exceeds adjusted ALARM HIGH limit
or falls below adjusted ALARM LOW limit.
Adult → Max: 220, Min: 15
Pediatric → Max: 220, Min: 15
Neonate → Max: 110, Min: 15
NIBP MAP LIMIT
MAP alarm is activated when the mean arterial pressure exceeds adjusted ALARM
HIGH limit or falls below adjusted ALARM LOW limit.(min:25 and max:200)
Adult → Max: 235, Min: 20
Pediatric → Max: 230, Min: 20
Neonate → Max: 125, Min: 20
AUTO/MANUAL/STAT
There are three modes of measurement available: MANUAL, AUTO and STAT. In the
MANUAL mode, only one measurement is performed. In the AUTO mode, measurement
is
repeated
1,2,3,5,10,15,20,30,45, 60, 90 minutes and 2, 4, 8, 12,16,20,24 hours. In STAT mode,
over
a
specified
period
of
time
10-9
User Manual
;
available
intervals
"symbol
are