Chapter 10: NIBP Monitoring

b) Technical alarms

Alarm
SELF-TEST
NIBP hardware module failure
FAILED
NIBP LOOSE
Cuff is completely unwrapped or
CUFF
no cuff attached.
Adult mode is used instead of
NIBP MODE
neonate mode (while neonate
ERROR
cuff is applied) or an occlusion
happens in air way
NIBP AIR
Air leak in cuff, tube or
LEAK
connector
NIBP AIR
Unstable pressure value (e.g.
PRESSURE
kinked hoses)
ERROR
Very weak patient signal due to
NIBP SIGNAL
a loosely wrapped cuff or
WEAK
extremely weak pulse from
patient.
NIBP ALARMS
Cause
10-13
Solution
Alarm level is set in NIBP
Window. By pressing ALARM
SILENCE, the message
background becomes gray and the
alarm is disabled and ignores this
fault.
Alarm level is set in NIBP
Window. By pressing ALARM
SILENCE, the message
background becomes gray and the
alarm is disabled and ignores this
fault.
Alarm level is set in NIBP
Window. By pressing ALARM
SILENCE, the message
background becomes gray and the
alarm is disabled and ignores this
fault.
Alarm level is set in NIBP
Window. By pressing ALARM
SILENCE, the message
background becomes gray and the
alarm is disabled and ignores this
fault.
Alarm level is set in NIBP
Window. By pressing ALARM
SILENCE, the message
background becomes gray and the
alarm is disabled and ignores this
fault.
Alarm level is set in NIBP
Window. By pressing ALARM
SILENCE, the message
background becomes gray and the
alarm is disabled and ignores this
fault.