Chapter 11: TEMP Monitoring
The TEMP WINDOW is as follows:
UNIT
Pick this item to set measurement unit. (options: ºC or ºF)
ALARM LEVEL
Selectable between 1 and 2. Level 1 represents the most serious case.
TEMP ALARM
Pick "ON" to enable TEMP alarm functions such as parameters blinking, audio alarm, and
light indicator.Pick "OFF" to disable the alarm functions and there will be a "
in the Parameter Area.
T1 LIMIT
T1 alarm is activated when the channel-1 temperature exceeds adjusted ALARM HIGH limit
or falls below adjusted ALARM LOW limit.(min:0.0 and max:50.0)
T2 LIMIT
T2 alarm is activated when the channel-2 temperature exceeds adjusted ALARM HIGH
limit or falls below adjusted ALARM LOW limit.(min:0.0 and max:50.0)
DT LIMIT
DT alarm is activated when the difference between channel-1 and channel-2 exceeds adjusted
ALARM HIGH limit or falls below adjusted ALARM LOW limit. (min:0.0 and max:50.0).
T1 is Channel-1 of temperature.
T2 is Channel-2 of temperature
DT is the temperature difference between the above two.