Chapter 14: GAS Monitoring (Sidestream)
NOTE:
For enabling ISA sensor, you can enter Gas window and set the monitor to
Measure mode.
GAS/RESP
Select this item to determine that respiration evaluation is performed by "Multi-gas"
or "RESP" module. Available options are "GAS" and "RESP". When selecting
"RESP", the system switches Multi-gas module to standby mode, and displays RESP
waveform and parameters.
FILL SIGNAL
Pick "ON" to show the waveform in filled form.
CO2 ALARM
Pick "CO2 ALARM" in GAS WINDOW to call up the following menu:
 CO2 ALARM
Pick "ON" to enable alarm functions such as parameters blinking, audio alarm and
light indicator. Pick "OFF" to disable the alarm functions and there will be a "
symbol in Multi-gas Parameter Area.
 ALARM LEVEL
Selectable between 1 and 2. Level 1 represents the most serious case.
 EtCO2 LIMIT
Alarm is activated when the EtCo2 exceeds adjusted ALARM HIGH or LOW limit
(Range: 0.4~13%V step 0.1%V)
Default for upper limit is 6.5%V and for lower limit is 2.6%V.
 FiCO2 HIGH
14-16
User Manual
"