Chapter 5: ECG Monitoring
LA or other leads are
ECG
not properly
CHECK LA
connected when ECG
OR ALL
lead is II for 3wire
lead set
RA or other leads are
ECG
not properly
CHECK RA
connected when ECG
OR ALL
lead is III for 3wire
lead set
ECG
C lead is not properly
CHECK C
connected to the
(C2, C3, C4,
patient.
C5, C6 )
After checking the mentioned solution, if the alarm persists, the ECG cable may be damaged and
you should contact with local After Sales Services.
Make sure that all electrodes and
patient cable are properly
connected
Make sure that all electrodes and
patient cable are properly
connected
Make sure that all electrodes esp.
C and ECG cable are properly
connected.
5-25
User Manual
Alarm level 2- the message is
displayed in yellow background. By
pressing ALARM SILENCE,
background becomes gray and the
alarm is disabled and ignores this
fault.
Alarm level 2- the message is
displayed in yellow background. By
pressing ALARM SILENCE,
background becomes gray and the
alarm is disabled and ignores this
fault.
Alarm level 2-the message is
displayed in yellow background. By
pressing ALARM SILENCE,
background becomes gray and the
alarm is disabled and ignores this
fault.