Chapter 9: SpO2 & Rainbow Parameters Monitoring
Warning
Do not place the accessories in any position that might cause it to fall on the patient.
Warning
Do not immerse sensor and patient cable completely in water, solvents, or cleaning
solutions because the sensor and patient cable are not waterproof.
Warning
ESU wire and SpO2 cable must not be tangled up.
Warning
Do not use the sensor on extremities with arterial catheter or venous syringe.
Warning
Do not start or operate the pulse co-oximeter unless the setup was verified to be correct.
Warning
Verify sensor cable fault detection before monitoring. Unplug the SpO2 sensor cable from
its socket, the screen will display the error message "
Warning
Do not repair or modify the pulse co-oximeter accessories. Injury to user or equipment
damage could occur. Contact with After Sales Services for servicing if necessary.
Changes or modifications shall void the guaranty for the pulse co-oximeter accessories.
Warning
Explosion hazard: Do not use the pulse co-oximeter in the presence of flammable
anesthetics or other flammable substance in combination with air, oxygen-enriched
environments, or nitrous oxide.
Warning
To protect from electric shock, always remove the sensor and completely disconnect the
pulse co-oximeter before bathing the patient.
Warning
If any measurement seems questionable, first check the patient's vital signs by alternate
means and then check the pulse co-oximeter for proper functioning.
SpO2 NO PROBE
9-8
User Manual
"