Chapter 12: IBP Monitoring

12.1 GENERAL

Specification:
Displaying and measuring range (for all labels)
Alarm ranges
Resolution
Accuracy
IBP stands for Invasive Blood Pressure. Patient Monitor measures direct blood pressure
(Systolic, Diastolic and Mean) of the selected blood vessel through two channels, and displays
two IBP waveforms.
Warning
The operator should avoid contacting with the conductive parts of the system when being
applied.
Warning
When using ESU (Electrosurgery equipment), the transducer and the cables should not
contact with the conductive part of ESU to protect patient against burns.
Warning
Disposable IBP transducer or domes should not be reused.
Warning
Before using dome make sure that its package is safe and check its expiry date.
Warning
Use only the pressure transducers listed in the
The specified transducers are designed to have the special ability to protect patient against the
electrical shock (especially for the leak current allowed), and it is protected against the effects of
a discharge of a cardiac defibrillator. It can be used in the surgical operation. During
defibrillation, the IBP waveform may be distorted temporarily.
IBP
ART
LVP
PAP
RVP
CVP
LAP
RAP
ICP
1 (mmHg)
+2 % or 2mmHg each one is greater
Accessories Chapter
12-2
User Manual
-50~300(mmHg)
-50~300(mmHg)
-50~300(mmHg)
-50~300(mmHg)
-50~120(mmHg)
-50~100(mmHg)
-50~100(mmHg)
-50~100(mmHg)
-50~100(mmHg)
-40~100(mmHg)
.