Download Print this page

Recorder - Saadat ALBORZ B9 Operator's Manual

Patient monitor

Advertisement

Chapter 18: Recording

18.2 RECORDER

Choose "RECORDER" in HOME WINDOW menu to call up the following window:
INTERNAL RECORDER
Pick "ON" to record via internal recorder.
Pick "OFF" to disable the internal recorder and record via the Central system.
TRACE1
To choose the first trace of printout record in manual recording.
Available options are ECG, SpO2, IBP1, IBP2, RESP, GAS and OFF.
TRACE2
To choose the second trace of printout record in manual recording.
Available options are ECG, SpO2, IBP1, IBP2, RESP, GAS and OFF.
NOTE:
You cannot choose same signal for TRACE 1 and TRACE 2.
TRACE3
To choose the third trace of print out record in manual recording.
Available options are ECG (different leads) and OFF.
Depending on the selected ECG CABLE TYPE in ECG menu, different options will be available
for TRACE 3.
-
If ECG CABLE TYPE is set to 3 WIRE, only OFF will be available for TRACE 3.
-
If ECG CABLE TYPE is set to 5 WIRE, ECG-I, ECG-II, ECG -III, ECG-V, ECG-
aVR, ECG-aVF, ECG-aVL and OFF will be available.