Chapter 9: SpO2 & Rainbow Parameters Monitoring
SPO2 SENSOR
The sensor connection to the
CHECK
system is not correct
CONNECTION
SPO2 LOW
SpO2 measurement does not
SIGNAL IQ
have confidence due to poor
signal quality caused by
excessive motion or other signal
interference.
SPO2 LOW PR
Pulse rate measurement does
CONFIDENCE
not have confidence due to poor
signal quality caused by
excessive motion or other signal
interference.
SPO2 LOW PI
PI measurement does not have
CONFIDENCE
confidence due to poor signal
quality caused by excessive
motion or other signal
interference.
PVI measurement does not have
confidence due to poor signal
SPO2 LOW PVI
quality caused by excessive
CONFIDENCE
motion or other signal
interference.
SPO2 LOW
SpOC measurement does not
SPOC
have confidence due to poor
CONFIDENCE
signal quality caused by
excessive motion or other signal
interference.
SPO2 LOW
SpCO SpO2 measurement does
SPCO
not have confidence due to poor
CONFIDENCE
signal quality caused by
Check the sensor
connection and, if
necessary, replace the
sensor and/or cable.
1-Assess the patient.
2-Check the sensor and
ensure proper sensor
application.
3-Change the sensor
site.
1-Assess the patient.
2-Check the sensor and
ensure proper sensor
application.
3-Change the sensor
site.
1-Assess the patient.
2-Check the sensor and
ensure proper sensor
application.
3-Change the sensor
site.
1-Assess the patient.
2-Check the sensor and
ensure proper sensor
application.
3-Change the sensor
site.
1-Assess the patient.
2-Check the sensor and
ensure proper sensor
application.
3-Change the sensor
site.
1-Assess the patient.
2-Check the sensor and
ensure proper sensor
9-23
User Manual
Alarm level 2- the message is
displayed in yellow background.
By pressing ALARM SILENCE,
background becomes gray and the
alarm is disabled for 120 sec. The
alarm is activated when SPO2
ALARM is "ON".
Alarm level 3- the message is
displayed in cyan background. By
pressing ALARM SILENCE,
background becomes gray and the
alarm is disabled and ignores this
fault.
Alarm level 3- the message is
displayed in cyan background. By
pressing ALARM SILENCE,
background becomes gray and the
alarm is disabled and ignores this
fault.
Alarm level 3- the message is
displayed in cyan background. By
pressing ALARM SILENCE,
background becomes gray and the
alarm is disabled and ignores this
fault.
Alarm level 3- the message is
displayed in cyan background. By
pressing ALARM SILENCE,
background becomes gray and the
alarm is disabled and ignores this
fault.
Alarm level 3- the message is
displayed in cyan background. By
pressing ALARM SILENCE,
background becomes gray and the
alarm is disabled and ignores this
fault.
Alarm level 3- the message is
displayed in cyan background. By
pressing ALARM SILENCE,