Download Print this page

Saadat ALBORZ B9 Operator's Manual page 144

Patient monitor

Advertisement

Chapter 12: IBP Monitoring
IBP1/IBP2/IBP3/IBP4 UNIT
Pick this item to adjust measurement unit. (Options: mmHg, Kpa, cmH2O)
Note:
To access PUMP PAGE in open heart surgery, select HOME > SETUP WINDOW >MAIN
DISPLAY >PUMP PAGE
For more information, see different page configurations in the configuration chapter.
Note:
IBP3 and IBP4 are only active in the ALBORZ B9 system.
IBP1/IBP2/IBP3/IBP4 LABEL
Suitable label should be selected, regarding the place of measurement. The available
pressure labels are:
Warning
IBP algorithm will vary according to the selected label. Therefore in the case of improper
label selecting, the accuracy of the measurement may be decreased.
Label
Definition
IBP
Invasive Blood Pressure
ART
Arterial Blood Pressure
LVP
Left Ventricle Pressure
PAP
Pulmonary Artery Pressure
RVP
Right Ventricle Pressure
CVP
Central Venous Pressure
LAP
Left Atrium Pressure
RAP
Right Atrium Pressure
ICP
Intracranial Pressure
12-5
User Manual

Advertisement

loading