Chapter 14: GAS Monitoring (Sidestream)
 FiN2O LIMIT
Alarm is activated when the FiN2O exceeds adjusted ALARM HIGH or LOW limit.
(Range: 1~82%V, step1%V)
Default for upper limit is 75%V and for lower limit is 35%V.
 EtAA LIMIT
Alarm is activated when the EtAA exceeds adjusted ALARM HIGH or LOW limit.
 FiAA LIMIT
Alarm is activated when the FiAA exceeds adjusted ALARM HIGH or LOW limit.
NOTE:
Alarm range and alarm limit default of different anesthesia agents are
mentioned in the table below:
Anesthesia agent
HAL
DES
ISO
SEV
ENF
 EtO2 LIMIT
Alarm is activated when the EtO2 exceeds adjusted ALARM HIGH or LOW limit.
(Range: 18~105%V step1%V)
Default for upper limit is 100% and for lower limit is 50%.
 FiO2 LIMIT
Alarm is activated when the FiO2 exceeds adjusted ALARM HIGH or LOW limit.
(Range: 18~105%V step 1%V)
Default for upper limit is 100%V and for lower limit is 50%V.
Alarm range
Step
0.1~5%
0.1%
0.1~18%
0.1%
0.1~5%
0.1%
0.1~8%
0.1%
0.1~5%
0.1%
14-18
User Manual
Alarm limit default
0.5~1.5%
5~10%
0.8~2%
1~3%
0.5~1.5%