Chapter 9: SpO2 & Rainbow Parameters Monitoring
The SpO2 WINDOW is as follows:
AVERAGE TIME
Available options are 2-4, 4-6, 8, 10, 12, 14 and 16.
SPO2 PLETH SWEEP
Available options for SpO2 PLETH SWEEP are 12.5 mm/s and 25 mm/s.
ALARM LEVEL
Selectable between 1 and 2. Level 1 represents the most serious case.
SpO2 ALARM
Pick "ON" to enable SpO2 alarm functions such as parameters blinking, audio alarm, and
light indicator. Pick "OFF" to disable the alarm functions and there will be a"
in the Parameter Area.
SpO2 LIMIT
SpO2 alarm is activated when the SpO2 exceeds adjusted ALARM HIGH limit or falls
below adjusted ALARM LOW limit. (min: 1 and max: 100, by step 1)
SpO2 SENSITIVITY MODE
It is only applicable to Masimo module. Available options for SpO2 SENSITIVITY are
NORMAL, MAX and APOD.
NORMAL: The perfusion threshold has different limits as the perfusion calculation is
data dependent .Specially; there is an intelligent algorithm which adjusts the low
perfusion limit in accordance with the quality of the incoming plethysmography
waveform between 0.5% and 0.02%. This mode provides the best combination of
sensitivity and probe-off detection performance. This mode is recommended for the
majority of patients.
9-15
User Manual