Chapter 18: Recording
-
If ECG CABLE TYPE is set to 10 WIRE, ECG-I, ECG-II, ECG -III, ECG-V, ECG-
aVR, ECG-aVF , ECG-aVL , V1, V2, V3, V4, V5, V6 and OFF will be available.
If available options for TRACE 3 are not suitable for the selected CABLE TYPE, the system will
automatically set TRACE 3 to OFF in 3WIRE mode and to ECG-II in 5 WIRE mode.
NOTE:
When TRACE 3 is active, the maximum gain option will be 0.5 mV.
RECORDER SWEEP
Available options for RECORDER SWEEP are 6, 12.5 and 25 mm/s.
MANUAL RECORD TIME
Available options for MANUAL RECORD TIME are "MANUAL", "10 sec", "20 sec",
"30 sec" and "CONTINUOUS".
PERIODIC TRACE 1
To choose the first channel trace of printout record in automatic recording.
Available options are ECG, SpO2, IBP1, IBP2, RESP, GAS and OFF.
PERIODIC TRACE2
To choose the second channel trace of printout record in automatic recording.
Available options are ECG, SpO2, IBP1, IBP2, RESP, GAS and OFF.
PERIODIC INTERVAL
To choose time interval in periodic recording. Available selections are 15min,
30min,1h,2h,4h,8h,12h,24h and OFF.
MANUAL REC DELAY
To create delay in manual recording. For example if you set this item to 10 seconds and
press Record button, data of the latest 10 seconds will be recorded.
Available options are 5 sec to 15 sec by step of 1 sec.
ALARM RECORD
If alarm recording for each parameter is set ON, it automatically starts recording when
alarms happen.
18-4