Chapter 7: ST Monitoring
7.2 ST ANALYSIS WINDOW
Pick "ST ANALYSIS "in the ECG WINDOW to call up the following menu:
ST ANALYSIS
Pick this item to enable or disable ST monitoring. The default is OFF. When the ST
monitoring is disabled "ST OFF" is displayed in ECG parameter area.
ST ALARM
Pick "ON" to enable alarm functions such as parameters blinking, audio alarm and light
indicator.Pick "OFF" to disable the alarm functions and there will be a "
the Parameter Area.
ALARM LEVEL
Selectable between 1 and 2. Level 1 represents the most serious case.
ST LIMIT
ST alarm is activated when the ST segment value exceeds adjusted ALARM HIGH limit
or falls below adjusted ALARM LOW limit. (Range: -2.0 ~ +2.0
Default for upper limit is +0.2 and for lower limit is -0.2.
EVENT DURATION
Pick this item to determine the time that a potential ST alarm condition must persist on
ECG waveform before the monitor classifies it as a valid alarm condition.
Available options for EVENT DURATION are OFF, 15s, 30s, 45s and 60s.The default is
OFF and alarm will be activated immediately if alarm condition happens.
ST REALERN
Pick to start a learning procedure. The message "RELEARN" is displayed in the message
area. The procedure will take about 20 seconds.
ECG/ ST ANALYSIS WINDOW
7-4
User Manual
" symbol in
step 0.1)