Saadat ALBORZ B9 Operator's Manual page 286

Patient monitor

Advertisement

Chapter 25: Trouble shooting
Fault Symptoms
Invalid SpO2 value
NIBP cuff could not inflate
NIBP measurement is not
successful
Invalid NIBP value
Invalid IBP value
Noisy IBP signal
After the catheter is inserted
into the patient body, the
message "ready for
measurement" does not
appear and the message
"Noisy baseline" remains on
the screen.
Inaccurate C.O value
Some advices to reduce measurement errors:
● NIBP
When NIBP measurement is taken, it is an important factor to set the measurement unit on
mmHg and connect the pressure cuff to the patient properly and according to instructions of
this manual. The most likely reason that the system doesn't display NIBP value is cuff failure
or leakage, therefore when dealing with this problem, use an intact cuff to test the system and
check air hose connection and other connections. If the problem is not removed, contact the
manufacturer's after sale service.
Possible Cause
●Patient movement during measurement
●Probe is placed in an unsuitable position.
●etc
●Incorrect air hose connection.
●Air hose occluded or tangled.
●Air hose or cuff leakage
●etc
●No cuff or Air hose is connected
●Wrong cuff placement
●Patient movement during measurement
●Low battery charge
●etc
●No zeroing before use
●Noisy source exists nearby the system or
accessories
●Faulty sensor
●etc
●The catheter is not placed in proper
position.
There is other noise source, for example
Electrocautery
● The manufacturer recommended
accessories are not used.
● Catheter type is not selected properly in
C.O Setup menu.
Injectate solution temperature is not zero (the
range of -5 to 5 0 °C)
●Calm patient
●Change the place of probe
●Call for service
●Check connection
●Check Air hose
●Change faulty accessory
●Call for service
●Check cuff and air hose
●Change cuff position
●Calm patient
●Connect the system to the mains power.
●Call for service
●Perform zeroing
●Keep system and cable away from noise
source
●Change the sensor
●Call for service
● Make sure that the catheter is inserted
properly.
Turn off the device caused noise and then
use C.O measuring module.
● Use the manufacturer recommended
accessories.
● Select the catheter type correctly in
Setup menu.
Make sure that the temperature of
injectate solution is zero.
25-2
User Manual
Correct Action