Chapter 24: Care and Cleaning
24.2 Cleaning and Disinfection
 General Points
Use only the substances approved by us and methods listed in this chapter to clean or disinfect
your equipment.
Manufacturer makes no claims regarding the efficacy of the listed chemicals or methods as a
means for controlling infection. For the method to control infection, consult your hospital's
Infection Control Officer or Epidemiologist. See also any local policies that apply within your
hospital.
Warning
1) Before cleaning the monitor or the sensors, make sure that the equipment is switched off
and disconnected from the power line.
2) Sterilization may cause damage to the device and is therefore not recommended for this
patient monitor otherwise indicated in the instructions delivered with accessories or your
hospital's servicing schedule.
3) If you see any signs of damage or deterioration in the device and its accessories, do not
use it, and if necessary, contact the after-sales service company.
4) Allow the monitoring system to dry completely before making connections. And please
make sure all connectors tightly connected to the system before using the system.
Please pay special attention to the following items:
1.The Patient Monitor and its belongings shall be kept dust-free.
2. Do not use strong solvents such as acetone or ammonia.
3. Most cleaning agents must be diluted before use.
4. Don't use rough or sharp material or your fingernail to remove stubborn stains.
5. Do not let the cleaning agent enter into the chassis of the system.
6. Do not leave the cleaning agents on any part of the equipment.
Warning
Do not use ETO gas to disinfect the monitor.
24-3