Chapter 12: IBP Monitoring
NOTE:
The alarm High/Low limits for SYS, DIA and MEAN of all labels are listed below.
Note that the CVP, LAP, RAP and ICP only have MEAN pressure, therefore the
alarm limits are only for MEAN. The alarm is enabled when the value exceeds the
adjusted limits.
Label
IBP
ART
LVP
PAP
RVP
CVP
LAP
RAP
ICP
IBP SCALE
Pick "IBP SCALE" in IBP WINDOW to call up the following menu:
The waveform and corresponding scale appears in the IBP waveform area with 3 dotted lines
representing HIGH limit scale, SIGN cursor, and LOW limit scale from the top to the bottom
that values of the three scales can be manually set or automatically by Auto scale. You can
change the high and low scales for IBP, ART and LVP labels by step of 10 and for PAP, RVP,
CVP, ICP, LAP and RAP labels by step of 5. The cursor changes by step of 1 for all labels.
Min Alarm Limit
(mmHg)
-50
-50
-50
-50
-50
-50
-50
-50
- 40
12-8
Max Alarm Limit
(mmHg)
300
300
300
120
100
100
100
100
100
User Manual
Step
(mmHg)
5
5
5
1
1
1
1
1
1