Saadat ALBORZ B9 Operator's Manual page 113

Patient monitor

Advertisement

Chapter 9: SpO2 & Rainbow Parameters Monitoring
SPO2 NO
SpO2 Sensor is not fully inserted
SENSOR
into the connector.
SPO2
SpO2 sensor has used all its
REPLACE
available monitoring time.
SENSOR
1. The SpO2 sensor is damaged
SPO2 SENSOR
2. SpO2 sensor is not
DEFECT
compatible.
1-SpO2 Sensor may be detached
from the patient.
SPO2 SENSOR
2-Sensor not connected to
OFF
patient properly.
3-Sensor is damaged.
When a single-patient-use sensor
SPO2 NO AD
is used, the adhesive portion of
SENSOR
the sensor is not connected.
message is displayed
again, replace cable.
Make sure that SpO2
sensor is correctly
connected into the
patient cable connector.
Replace the SpO2
sensor.
1. Make sure that SpO2
sensor is properly
attached to the cable
connector.
2. Restore power to the
instrument. If this
message is displayed
again, replace sensor.
1-Disconnect and
reconnect sensor.
Reattach sensor.
2-Properly reapply the
sensor on the patient
and reconnect the
sensor to the monitor or
patient cable.
3-Replace the sensor.
Ensure the adhesive
portion is firmly
connected to the sensor.
9-21
User Manual
Alarm level 3- the message is
displayed in cyan background. By
pressing ALARM SILENCE,
background becomes gray and the
alarm is disabled and ignores this
fault.
Alarm level 2- the message is
displayed in yellow background.
By pressing ALARM SILENCE,
background becomes gray and the
alarm is disabled and ignores this
fault.
Alarm level 2- the message is
displayed in yellow background.
By pressing ALARM SILENCE,
background becomes gray and the
alarm is disabled and ignores this
fault.
Alarm level 2- the message is
displayed in yellow background.
By pressing ALARM SILENCE,
background becomes gray and the
alarm is disabled and ignores this
fault.
Alarm level 3- the message is
displayed in cyan background. By
pressing ALARM SILENCE,