Chapter 6: Arrhythmia Monitoring
RATE
With count, you can determine the point at which an event call is triggered.
You cannot modify the rate for "ASYSTOLE", "VFIB", "COUPLET", "BIGEMINY",
"TRIGEMINY", "PAUS", " AFIB" and "FREQUENT PVCs" .
"RUN" and "AIVR" derive their rate settings from "VTAC" and cannot be modified.
COUNT
With rate, you can determine the point at which an event call is triggered.
You can't modify the count for "ASYSTOLE", "VFIB", "COUPLET", "BIGEMINY",
"TRIGEMINY","TACHY","BRADY", "AFIB" and "PAUS".
Count of "AIVR" is ≥3 and cannot be modified.
ARCHIVE
You can determine whether the selected event is stored, recorded automatically or both.
You can view stored events on ARR EVENT RECALL Window.
STR: Stores selected arrhythmia event.
REC: Automatically generates a recording of selected event.
STR/REC: Event is stored and recorded simultaneously.
OFF: No action if arrhythmia event activates.
ALL ALM LEVEL
Press to set the level of all arrhythmia alarms to the same value or to disable all of them.
ALL ARCHIVE
Press to set all arrhythmia ARCHIVE condition to the same state.
Arrhythmia event
VTAC
RUN
AIVR
TACHY
BRADY
Arrhythmia event
VTAC
RUN
FREQUENT PVCs
Rate setting
100-200 step by 10
Same as VTAC rate
<VTAC rate-1
100-200 step by 10
30-105
step by 5
Count setting
5-12
step by 1
(VTAC count -1) ~3 step by
1
1-15
step by 1
6-7
User Manual