Chapter 11: TEMP Monitoring
NOTE:
Please be noted that the metal side of probe should be used for making measurements.
WARNING
Using ESU with temperature measurement simultaneously may cause patient burn. If
possible, remove the probe from patient contact before activating the surgical unit or other
RF source. If probe must be used simultaneously with electrosurgical apparatus, hazards
can be reduced by selecting a temperature monitoring point which is remote from the
expected RF current path to the ground return pad.
WARNING
Over straining will result in mechanical damage of the probes.
WARNING
The temp probes should be calibrated every two years or according to hospital calibration
schedule. Contact After Sale Service to perform probe calibration.
WARNING
The temperature probes carry a one-year warranty on workmanship, components and
accuracy tolerances. Probe life with normal use should exceed one year.
11.2 TEMP WINDOW
The following items are displayed in TEMP parameter window.
TEMP PARAMETER WINDOW
11-3