Chapter 8: RESP Monitoring
RR ALARM
Pick "ON" to enable RESP alarm functions such as parameters blinking, audio alarm and
light indicator. Pick "OFF" to disable the alarm functions, and there will be a"
in the Parameter Area.
RR LIMIT
RESP alarm is activated when the respiration rate exceeds adjusted ALARM HIGH limit or
falls below adjusted ALARM LOW limit. (min:5 and max:150)
APNEA LIMIT
To set the standard of judging an apnea case. It can be set to 10 – 40 seconds and OFF and
increases/decreases by 10s. When you select OFF, the message "APPNEA LIMIT: OFF"
will appear at the bottom of RR window in red color.
NOTE:
APNEA alarm is always enabled with level 1 and ON/OFF status of RR ALARM
has not any effect on it.
CAPNO/RESP
To select "RESP" or "CAPNO" module for measuring respiratory rates. Available options
are "RESP" and "CAPNO". In "RESP" mode, CAPNO module is set to standby mode and
RESP parameters and waveform are displayed.
ALARM RECORD
See the chapter "RECORDING".