Chapter 12: IBP Monitoring
Warning
When using PUMP PAGE, the IBP label must be set to CVP.
IBP SWEEP
Available options for IBP SWEEP are 3,6,12.5 and 25mm/s.
IBP GRID
Select "ON" to divide each IBP signal to 5 parts with white dot lines.
IBP FILTER
In order to have a more clear and detailed waveform, 3 filter types can be selected:
Available options are 22Hz, 16Hz, and 8Hz.
22Hz : Recommended in normal use and the most clinical situation. It has the most
measuring accuracy among the called filters.
16Hz : When the signal is a bit noisy.
8Hz
: This mode is recommended to reduce noise and interface resulted from ESU and
also when the system has a high noise level or doesn't have equipotential earth. While
using this filter the measuring accuracy might be decreased.
ADJUST SCALE
Pick < ADJUST SCALE> in IBP WINDOW to adjust the scale automatically. The scales
are adjusted in a way that IBP signal occupies minimum 80% of IBP waveform area.
ALWAYS AUTO SCALE
This item will be available in IBP WINDOW if you set AUTO SCALE to ON. Set ON this
item to adjust the scale automatically. In this condition High, Low and Sign scales are not
displayed. This item is applicable for when there are large pressure variations and IBP signal
exceeds the selected scale. Set OFF this item to see scale values (High, Low and Sign). In
this condition you can adjust scale manually or using ADJUST SCALE.
NOTE:
When "ALWAYS AUTO SCALE" is set ON:
-
Scale values (High, Low and Sign) are not displayed.
-
Sign scale is displayed in the area between High and Low scales.
-
AUTO SCALE and IBP SCALE become inactive.
Set OFF this item to terminate above condition.
12-6