Chapter 10: NIBP Monitoring
Error occurs in pump, A/D
SYSTEM
sampling, pressure transducer or
FAILURE
software.
NIBP NO
No NIBP module is installed.
MODULE
NIBP LOW
The battery charge is not enough
BATTERY
to measure NIBP.
NIBP MODULE
Some errors occur during
ERROR
measurement.
Alarm level of above messages is set in NIBP WINDOW. By pressing ALARM SILENCE, alarm
will be disabled and ignores this fault and message background will change to gray.

c) Messages

Message
NIBP stop key is pressed
NIBP STOP
PRESSED
during measurement.
NIBP
Successful leakage test
LEAKAGE O.K
NIBP MESSAGES
Cause
10-15
Alarm level is set in NIBP
Window. By pressing ALARM
SILENCE, the message
background becomes gray and the
alarm is disabled and ignores this
fault.
Alarm level is set in NIBP
Window. By pressing ALARM
SILENCE, the message
background becomes gray and the
alarm is disabled and ignores this
fault.
Alarm level is set in NIBP
ALARM MENU. By pressing
ALARM SILENCE, the message
background becomes gray and the
alarm is disabled and ignores this
fault.
Alarm level is set in NIBP
ALARM MENU. By pressing
ALARM SILENCE, the message
background becomes gray and the
alarm is disabled and ignores this
fault.
Solution
User Manual
Explanation