Chapter 2: Configuration
IBP) (MAIN DISPLAY= PAGE 6 mode)
*
You can monitor ECG, SpO2, IBP1, IBP2 and RESP signals as well as numeric
P7
:
parameters HR, PVCs, ST, SpO2%, PR, IBP1,IBP2, NIBP,TEMP and RESP/CO2 (MAIN
DISPLAY= PUMP PAGE mode).
NOTE:
To access PUMP Page, select SETUP from Home menu and then set MAIN
DISPLAY to PUMP Page.
*
You can monitor ECG signal (12 traces) and numeric parameters of P1.
P8
:
(Only with settings of CABLE TYPE: 10WIRES and DISPLAY FORMAT: 12 TRACES,
you can access this page).
NOTE:
To access P8, select ECG LEAD from ECG window and then set CABLE TYPE to
10WIRES and DISPLAY FORMAT to 12.
*
If CO2 is selected, ECG, SpO2, IBP2 and BFA signals as well as numeric parameters
P9
:
HR, SpO2, IBP2, T1, T2 and NIBP will be displayed (In this page only BFA settings are
available).
If RESP is selected, ECG, SpO2, IBP1, IBP2 and BFA signals as well as numeric parameters
HR, SpO2, IBP1,IBP2, T1, T2, NIBP and RESP will be displayed (In this page only BFA
settings are available).
NOTE:
You can access P9 by clicking on BFA window (in all pages except Page 6).
PUMP PAGE
PUMP page is an operating mode that can be selected via SETUP menu. This page provides the
following conditions depending on events occurring for ECG and IBP modules during open heart
operation.
1- The word "PUMP" is shown on ECG signal.
2- If ASYSTOLE condition occurs, the message "ECG ASYSTOLE" will be shown and audible
alarm with high level will be activated.
3-You can disable the alarm sound by pressing Alarm Silence key, but the message "ECG
ASYSTOLE" will remain on the screen.
4- If other alarm except ASYSTOLE alarm occurs, audible alarm will sound corresponding to
2-5