Chapter 14: GAS Monitoring (Sidestream)
Message
CO2 SENSOR
STANDBY MODE
ZEROING IN
PROGRESS
Indicator status on the ISA sensor:
Steady green light
Blinking green light
Steady blue light
Steady red light
Blinking red light
Cause
Manual setting and if no
breath is detected for 30 min
and ETCO2 is less than 4
mmHg for more than 30 min
or when the monitor does not
detect the sampling line.
The zeroing procedure is
.
being conducted
Solution
Enter GAS window and
set WORK MODE to
MEASURE.
ISA in operation and OK
Zeroing in progress
Anesthetic agent present
ISA sensor error
Check sampling line
14-24
User Manual
Explanation
After that the zeroing
procedure is completed, this
message and a flat line signal
will be displayed.