Chapter 9: SpO2 & Rainbow Parameters Monitoring
SpMet violates adjusted low
SpMet LOW
alarm limit
SpHb violates adjusted high
SpHb HIGH
alarm limit
SpHb violates adjusted low
SpHb LOW
alarm limit

b) Technical alarms

Alarm
SPO2 NO
SpO2 cable is not fully inserted
CABLE
to the patient monitor system.
SPO2
The life of the SpO2 cable has
REPLACE
expired.
CABLE
SPO2 CABLE
1. The SpO2 cable is damaged
DEFECT
2. SpO2 cable is not compatible.
SPO2 ALARMS
Cause
Make sure that the
SpO2 cable is correctly
connected into the
monitor.
Replace the SpO2
cable.
1. Make sure that the
Masimo SpO2 cable is
correctly connected
into the monitor.
2. Restore power to the
instrument. If this
9-20
● SpMet value blinks.
● The alarm indicator flashes
● The alarm message is
displayed in a background
corresponding to its level.
● SpHb value blinks.
● The alarm indicator flashes
● The alarm message is
displayed in a background
corresponding to its level.
● SpHb value blinks.
● The alarm indicator flashes
● The alarm message is
displayed in a background
corresponding to its level.
Solution
Alarm level 3- the message is
displayed in cyan background. By
pressing ALARM SILENCE,
background becomes gray and the
alarm is disabled and ignores this
fault.
Alarm level 2- the message is
displayed in yellow background.
By pressing ALARM SILENCE,
background becomes gray and the
alarm is disabled and ignores this
fault.
Alarm level 2- the message is
displayed in yellow background.
By pressing ALARM SILENCE,
background becomes gray and the
alarm is disabled and ignores this
fault.
User Manual
Activated
Activated
Activated
Explanation