Download Print this page

Saadat ALBORZ B9 Operator's Manual page 78

Patient monitor

Advertisement

Chapter 6: Arrhythmia Monitoring
Ventricular Trigeminy: Sequence of
TRIGEMINY
beats with the pattern : normal,
ARRHYTHMIA
normal, PVC, normal, normal, PVC
Sinus Tachycardia: HR ≥ TACHY rate
TACHY
setting. A PVC or other abnormal beat
ARRHYTHMIA
breaks the analysis sequence and
restarts analysis.
Sinus Bradycardia: HR≤ BRADY rate
BRADY
setting. A PVC or other abnormal beat
ARRHYTHMIA
breaks the analysis sequence and
restarts analysis.
AFIB
Atrial Fibrillation: Formation of QRS
ARRHYTHMIA
complexes in irregular intervals
PAUS
Actual R-R interval more than 2.1
ARRHYTHMIA
times of the average R-R interval.
More than N (event count set in the
FREQUENT
ARR SETUP WINDOW) PVC per
PVCs
minute.
●The alarm indicator flashes.
●The alarm message is displayed
in a background corresponding to
its level.
●The alarm indicator flashes.
●The alarm message is displayed
in a background corresponding to
its level.
●The alarm indicator flashes.
●The alarm message is displayed
in a background corresponding to
its level.
●The alarm indicator flashes.
●The alarm message is displayed
in a background corresponding to
its level.
●The alarm indicator flashes.
●The alarm message is displayed
in a background corresponding to
its level.
●The alarm indicator flashes.
●The alarm message is displayed
in a background corresponding to
its level.
6-12
User Manual
ON)
Activated
(If ARR MONITOR
and Alarm
Level are switched ON)
Activated
(If ARR MONITOR
and Alarm
Level are switched
ON)
Activated
(If ARR MONITOR
and Alarm
Level are switched ON)
Activated
(If ARR MONITOR
and Alarm
Level are switched
ON)
Activated
(If ARR MONITOR
and Alarm
Level are switched ON)
Activated
(If ARR MONITOR
and Alarm
Level are switched ON)

Advertisement