Download Print this page

Saadat ALBORZ B9 Operator's Manual page 76

Patient monitor

Advertisement

Chapter 6: Arrhythmia Monitoring
NOTE:
You can do relearn procedure by selecting <ST RELEARN> in ECG/ST
ANALYSIS window.
NOTE:
In most situations the learning phase takes about 20 seconds.
NOTE:
If the monitor couldn't find 6 matching beats after 20 seconds, the relearn
procedure continues and the "RELEARN" message remains on the display, till
acceptable condition happens.
NOTE:
While the monitor is in learning phase, all arrhythmia alarms and trend collection
are suspended.
NOTE:
Before starting learning procedure, verify the quality of the ECG signal and ensure
that the patient's ECG displays a normal reference pattern.
NOTE:
The monitor automatically begins to learn a reference template whenever you
execute any of the following tasks (If ARR ANALYSIS is ON and there is no
technical ECG alarm active, like CHECK LEAD):
-
Turning on the monitor
-
Connecting ECG cable.
-
Changing ECG lead configuration.
-
Choosing "NEW" in HOME / PATIENT INFORMATION
NOTE:
It is recommended to perform relearn procedure under the following conditions:
-
A lead is reconnected or electrodes are repositioned.
-
Eight hours have passed since last reference complex learned.
-
Other significant changes appear on the morphology of the patient's ECG.
6-10
User Manual

Advertisement

loading