Chapter 9: SpO2 & Rainbow Parameters Monitoring
When a single-patient-use sensor
SPO2
is used, the life of the adhesive
REPLACE AD
portion of the sensor has
SENSOR
expired.
When a single-patient-use sensor
SPO2 AD
is used:
SENSOR
1. The adhesive portion of the
DEFECT
sensor is damaged.
2. SpPO2 sensor is not proper.
This may be caused by excessive
SPO2
ambient light sources such as
AMBIENT
surgical lights or direct sunlight,
LIGHT
or other.
SPO2
RAINBOW
SpO2 hardware error
HARDWARE
FAIL
SPO2 PROBE
Failure to properly operate
DEFECT
sensor or cable or both of them.
Replace the adhesive
portion of the sensor.
1. Make sure that SpO2
sensor is properly
attached to the cable
connector.
2. Power off and then
on the system. If this
message is displayed
again, replace the
adhesive portion of the
sensor.
In the case of using
rainbow sensor, place a
Masimo Optical Light
Shield over the sensor.
Restore power to the
instrument. If this
message is displayed
again,contact After
sales service of
manufacturer.
Check the function of
the sensor and the cable
separately and replace
the defective part.
9-22
User Manual
background becomes gray and the
alarm is disabled and ignores this
fault.
Alarm level 2- the message is
displayed in yellow background.
By pressing ALARM SILENCE,
background becomes gray and the
alarm is disabled and ignores this
fault.
Alarm level 2- the message is
displayed in yellow background.
By pressing ALARM SILENCE,
background becomes gray and the
alarm is disabled and ignores this
fault.
Alarm level 2- the message is
displayed in yellow background.
By pressing ALARM SILENCE,
background becomes gray and the
alarm is disabled for 120 sec. The
alarm is activated when SpO2
ALARM is "ON".
Alarm level 2- the message is
displayed in yellow background.
By pressing ALARM SILENCE,
background becomes gray and the
alarm is disabled for 120 sec. The
alarm is activated when SPO2
ALARM is "ON".
Alarm level 2- the message is
displayed in yellow background.
By pressing ALARM SILENCE,
background becomes gray and the
alarm is disabled for 120 sec. The
alarm is activated when SPO2
ALARM is "ON".