Chapter 12: IBP Monitoring
NOTE:
Take the following actions for calibration if MEDEX transducer is used:
Pick <CALIB> in IBP WINDOW to access the window shown in Figure 9-7.
Set IBP1 and IBP2 to 100mmHg and push down Calib button of the transducer for
about 10 seconds.
SIGNAL SELECTION
This option is only available in the page that four channels of IBP are displayed. (This item only
active in Alborz B9)
Select "IBP1,2" to observe IBP1 and IBP2 signals and select "IBP3,4" to monitor IBP3 and
IBP4 signals.
ART CATH. DISCONNECT ALM
If catheter is disconnected from the patient during the pressure measurement, the following
conditions will occur:
The pressure drops dramatically.
IBP signal becomes static and the MEAN pressure falls below 10 mmHg.
The heart activity is not shown and the signal is displayed as a flat line.
In this condition, "IBP CATHETER DISCONNECT" alarm with level 1 will be enabled for
maximum 10 seconds.
To trigger the alarm, set label to ART or IBP and enable "ART CATH. DISCONNECT".
(a) IBP WINDOW
12-13
User Manual