Chapter 14: GAS Monitoring (Sidestream)
Alarm is activated when the FiCo2 exceeds adjusted ALARM HIGH limit (Range:
0.4~13 %V step 0.1%V). Default for upper limit is 1.3%V.
 AWRR LIMIT
Alarm is activated when the AWRR exceeds adjusted ALARM HIGH or LOW limit.
(Range: 1-120BrPM)
Default for upper limit:
Default for lower limit:
 APNEA LIMIT
Pick it to set the standard of judging an apnea case. It sets to 20 - 60 seconds and
"OFF" and increases/decreases by 5s.
GAS ALARM
Pick "GAS ALARM" in GAS WINDOW o call up the following menu:
 N2O ALARM, AA ALARM and O2 ALARM
Pick "ON" to enable alarm functions such as parameters blinking, audio alarm and
light indicator. Pick "OFF" to disable the alarm functions and there will be a "
symbol in Multi-gas Parameter Area.
 ALARM LEVEL
Selectable between 1 and 2. Level 1 represents the most serious case.
 EtN2O LIMIT
Alarm is activated when the EtN2O exceeds adjusted ALARM HIGH or LOW limit.
(Range: 1~100%V step1%V)
Default for upper limit is 75%V and for lower limit is 35%V.
Adult/Pediatric:
30BrPM
Neonate:
60BrPM
Adult/Pediatric:
5BrPM
Neonate:
15BrPM
14-17
User Manual
"