Chapter 5: ECG Monitoring
Warning
For patients with pacemaker, PACE DETECT must be switched "ON", otherwise, the pace
pulses may affect HR counting and result in low precision of HR value.
Warning
For the patients with pacemaker, the monitor may continue to count the pacemaker rate as
heart rate during the occurrence of cardiac arrest or some arrhythmias. Do not rely
entirely upon monitor alarms. Keep the patients with pacemaker under close surveillance.
ECG CALIB
Pick "ON" to view 1mV calibrated ECG wave. When it is "ON", the calibration waveform will
be displayed until closing the ECG WINDOW or turning the rotary switch.
ALARM
Pick "ALARM" in ECG WINDOW to call up the following menu:
ALARM LEVEL
Selectable between 1and 2. Level 1 represents the most serious case.
HR ALARM
Pick "ON" to enable alarm functions such as parameters blinking, audio alarm and light
indicator. Pick "OFF" to disable the alarm functions and there will be a "
the Parameter Area.
HR LIMIT
ECG alarm is activated when the heart rate exceeds adjusted ALARM HIGH value or
falls below adjusted ALARM LOW value (min:30 and max:250)
5-20
User Manual
" symbol in