Chapter 14: GAS Monitoring (Sidestream)
14.5 Indicator status on the ISA sensor:
Steady green light
Blinking green light
Steady blue light
Steady red light
Blinking red light

14.6 Pre-use check

Before connecting the Nomoline Family sampling line to the breathing circuit, do the following:
1. Connect the sampling line to the ISA gas inlet connector.
2. Check that the gas inlet connector is lit with a steady green light.
3. For ISA OR+:
Check that the O2 reading on the monitor is correct (21 vol%).
4. Breathe briefly into the sampling line and check that monitor displays a valid CO
waveform and valid values.
5. Occlude the sampling line with a fingertip and wait for 10 seconds.
6. Check that occlusion alarm is displayed on the monitor and that the gas inlet connector
shows a flashing red light.
7. If applicable:
Perform a tightness check of the patient circuit with the sampling line attached.
NOTE:
Variations in barometric pressure do not have any effects due to internal
barometric pressure compensation.
NOTE:
There are no adverse effects on stated performance due to cycling pressure of up to
10 KPa.
Warning
ISA in operation and OK
Zeroing in progress
Anesthetic agent present
ISA sensor error
Check sampling line
14-9