18.6 Recorder Alarm Messages

Message
Rec. Software Error
Recorder Fault
Rec Door Open
Rec Paper Out
Print head High Temp
Print head High Vol.
Print head Low Vol.
Time out Error
All alarm messages of the recorder are level 2. The messages are displayed in the yellow
background. By pressing Silence key, the message background will change to the gray and the
system will ignore this fault.
Cause
Software error
Hardware error
The recorder door is open
Recorder paper has been finished.
The thermal head is too hot.
Print head voltage is high.
Print head voltage is low.
The recorder could not record.
Solution
Turn the system off and then on .If the problem
persists, contact after sales service of manufacturer.
Turn the system off and then on. If the problem
persists, contact after sales service of manufacturer.
Close the recorder door.
Insert a new paper roll.
Stop operation for some minutes.
Turn the system off and then on. If the problem
persists, contact after sales service of manufacturer.
Turn the system off and then on. If the problem
persists, contact after sales service of manufacturer.
Turn the system off and then on. If the problem
persists, contact after sales service of manufacturer.