Chapter 14: GAS Monitoring (Sidestream)
NOTE:
Do not operate the ISA sidestream gas analyzer outside the specified operating
environment.
Warning
Measurements can be affected by mobile and portable RF communications equipment. It
should be assured Make sure that the ISA gas analyzer is used in the electromagnetic
environment specified in EMC section of this manual.
Warning
The ISA sidestream gas analyzers are not designed for MRI (magnetic resonance imaging)
environments. During (MRI) scanning, ISA must be placed outside the MIR suite.
Warning
Use of high frequency electrosurgical equipment in the vicinity of the ISA/monitor may
produce interference and cause incorrect measurements.
Warning
Do not use the Nomoline Airway Adapter Set Infant with adult/pediatric patients.
Warning
Do only use sample lines intended for anesthetic agents if N
being used.
Warning
Do not re-use disposable single-patient use Nomoline Family sampling lines due to the risk
of cross contamination.
Warning
Due to the risk of patient cross-infection, always use a bacteria filter on the exhaust port
side if sampled gas is intended to be re-breathed.
Warning
Exhaust gases should be returned to the patient circuit or to a scavenging system.
O and/or anesthetic agents are
2
14-8
User Manual