Chapter 10: NIBP Monitoring
MODULE CHECK
Select this item to open the respective menu after 5 seconds delay.Available options are
"NIBP MANOMETER","NIBP LEAKAGE","MODULE SELF TEST" and "MODULE
STOP".
NOTE:
Below tests must only be done by trained and authorized personnel.
■ NIBP MANOMETER
Wrap the cuff around a rigid cylinder. Connect a calibrated reference manometer and a ball
pump by means of a T-piece connector and hoses to the monitor. Set the monitor in" NIBP
MANOMETER" mode. Inflate the pneumatic system to 0, 50 and 200 mmHg by ball pump
separately. The difference between the indicated pressure of the reference manometer and the
indicated pressure of the monitor should not exceed ±3 mmHg.
■ NIBP LEAKAGE
Wrap the cuff around a cylinder of an appropriate size, and the circumference of the applied cuff
does not exceed that of the cylinder more than 7%.Set the monitor in "NIBP LEAKAGE" mode.
The monitor inflates the cuff up to 200mmHg and keeps it constant for 20 sec .If air leakage
result is satisfactory, "NIBP LEAKAGE OK" message is displayed; otherwise you will receive
"PNUMATIC LEAK" message.
■ NIBP SELFTEST
The general objective of the SELFTEST is to check the general status of the NIBP module,
including the function of sensors and valves.
10-11
User Manual