Download Print this page

Alarm Recall - Saadat ALBORZ B9 Operator's Manual

Patient monitor

Advertisement

ALARM RECALL

17.5 ALARM RECALL *
Choose "ALARM RECALL" in HOME WINDOW to access the following window in which you
can store and review data of alarm occurrence time.
Alarm data including the number of alarm event, date, time and HR, SpO2, IBP1-SYS/DIA
(MAP), IBP2-SYS/DIA (MAP), PR, T1, T2, ETCo2, FICo2 and NIBP values as well as page
number are displayed in this window.
The parameter value which triggered the alarm will be displayed in the red and the parameter value
of the previously triggered alarm which is still active will be displayed in the yellow color. The
parameter value of inactive alarm is displayed in the white color.
 Reviewing alarm list
Data of 20 alarm events are displayed in HOME/ALARM RECALL WINDOW. Up to seven alarm
events are displayed in each page of ALARM RECALL WINDOW. If there are more than seven
alarms, other pages will be available. By clicking on "UP-DOWN", you can access different pages
of HOME/ALARM RECALL WINDOW.
If a new alarm occurs after that Alarm Recall list is filled up, the last alarm event will be replaced
by the new alarm event and will be displayed at the top of the list.
 Selecting an alarm event
Click on ▲▼ to select an alarm event in the list.
 Deleting an alarm event
Select an alarm event and click on "DELETE" to delete the alarm.
17-9