Chapter 14: GAS Monitoring (Sidestream)

b) Technical alarms

Alarm
CO2 SYSTEM FAULT
#1,2,3,4
CHECK SAMPLING
LINE
SAMPLING LINE
CLOGGED
CO2 INVALID
O2 INVALID
N2O INVALID
AGENT INVALID
Cause
Sensor error
sampling line is not working
Sampling line occlusion
CO2 outside specified
accuracy range.
O2 outside specified
accuracy range.
N2O outside specified
accuracy range.
Agent outside specified
accuracy range.
Solution
Turn the system off and
on. If the problem still
exists, contact after sales
service of manufacturer.
Replace the sampling
line
Replace the sampling
line
Zero the system. If the
problem still exists,
contact after sales service
of manufacturer.
Zero the sensor, if the
problem still exists,
contact After sales
service of manufacturer.
Zero the sensor. If the
problem still exists,
contact After sales
service of manufacturer.
Zero the sensor. If the
problem still exists,
contact After sales
service of manufacturer.
14-22
User Manual
Explanation
Alarm level 2- the message is
displayed in yellow
background. By pressing
ALARM SILENCE,
background becomes gray and
alarm is disabled and ignores
this fault.
Alarm level 2- the message is
displayed in yellow
background. By pressing
ALARM SILENCE,
background becomes gray and
the alarm is disabled for 120
sec
Alarm level 2- the message is
displayed in yellow
background. By pressing
ALARM SILENCE,
background becomes gray and
alarm is disabled for 120 sec.
Alarm level 2- the message is
displayed in yellow
background. By pressing
ALARM SILENCE,
background becomes gray and
alarm is disabled and ignores
this fault.
Alarm level 2- the message is
displayed in yellow
background. By pressing
ALARM SILENCE,
background becomes gray and
alarm is disabled and ignores
this fault.
Alarm level 2- the message is
displayed in yellow
background. By pressing
ALARM SILENCE,
background becomes gray and
alarm is disabled and ignores
this fault.
Alarm level 2- the message is
displayed in yellow
background. By pressing
ALARM SILENCE,
background becomes gray and
alarm is disabled and ignores
this fault.