Chapter 12: IBP Monitoring
ALARM
Pick "IBP ALARM" in IBP WINDOW to call up the following menu:
IBP1/IBP2/IBP3/IBP4 ALARM
Pick "ON" to enable alarm functions such as parameters blinking, audio alarm and light
indicator. Pick "OFF" to disable the alarm functions and there will be a "
Parameter Area.
IBP1/IBP2/IBP3/IBP4 ALARM LEVEL
Selectable between 1and 2. Level 1 represents the most serious case.
IBP ALARM RECORD
See the chapter "RECORDER".
SYS ALM
SYS alarm is activated when the systolic pressure exceeds adjusted ALARM HIGH limit or
falls below adjusted ALARM LOW limit.
DIA ALM
DIA alarm is activated when the diastolic pressure exceeds adjusted ALARM HIGH limit or
falls below adjusted ALARM LOW limit.
MEAN ALM
MEAN alarm is activated when the mean pressure exceeds adjusted ALARM HIGH limit or
falls below adjusted ALARM LOW limit.
12-7