In each curve, the vertical axis represents the numerical value of the parameter and the
horizontal axis of the time. it is possible to change time frame by clicking on the written time
in the middle of horizontal axis area. These time frames are as follows 60 min, 6 h, 12 h, 24
h, 36 h and 72 h. It should be noted that the changing of the time frame applied on a curve is
also applied simultaneously on two other curves. it is also possible to change the scale of
parameter by clicking on the vertical axis.
The parameters that could be trended are listed in below table:
Basic/ Fundamental
MV Spont
Rate Spont
5-7

Alarms

Selecting Alarms will load Alarms Menu in which you can set alarm limits. There is also an
Event log option on this page, which is described following. Set alarm limits in two ways:
1. Click on alarm limits (high and low) and set your desired value.
2. Select AutoSet key to set alarm limits of all parameters except Ppeak, Leak, Apnea
time, Vt lim automatically based on setting values which are listed below:
Figure 5-9 Trend
Vti
Vte
MV
Rate
P peak
PendInsp
P mean
PFe
Advanced
Pplateau
PEEP
Auto PEEP
Ti
Te
Ti/T tot
Leak
O
2
Mechanics
R insp
R exp
C stat
C dyn
WOBimp
RSBI
RCe
Gas Temp
34