Chapter 22:Technical Specifications

CLASSIFICATION

Protection against electroshock
Protection
Mode of operation
Harmful Liquid Proof Degree
Method of disinfection
Safety of anesthetic mixture

General

Display
Alborz B9
Waveforms
Numeric Parameters
Operation Method
AC Power

ECG

Lead & Wire Options
Selectable: 3 ,5 or 12 Leads
3 ECG Leads I, II, III
5 Leads ECG: I,II,III,V,aVR,aVF,aVL
12 Leads ECG : I,II,III, aVR,aVF,aVL , V1,V2, V3, V4, V5. V6
Dynamic Range
Lead Off Current
Gain
Calibration
Filters
CMRR
Internal Noise
Input Impedance
QRS Detection
Heart Rate Range
Accuracy
Tall T-Wave
Pacer Detection/Rejection
Class I, Type CF for all modules (except Multi-gas, NIBP and BFA
modules that are BF) (based on IEC 60601-1).
Against Electro surgery and Defibrillator (Except BFA)
Continues operation equipment
IPX1
Refer to each module's chapters and chapter Care & Cleaning for detail.
Not suitable for use in the presence of a flammable anaesthetic mixture
with air or with oxygen or nitrous oxide.
1366  768
TFT/LED
COLOR
ECG, SpO2, IBP1, IBP2, RESP/GAS,EEG (Freezable), C.O.
HR,PVCs,ST,SpO2, PR, NIBP (SYS, DIA, MAP),
IBP1(SYS,DIA,MAP), IBP2(SYS,DIA,MAP), RR, T1, T2, DT, EtCo2,
FiCo2, AWRR, EtN2O, FiN2O, EtO2, FiO2, EtAA, FiAA, BFI, BS%,
EMG%, SQI%, C.O, Alarm Limits.
Membrane/Keys and rotary knob, Touch Screen
100 - 240 VAC, 50/60 Hz , Ip: 0.9 – 0.4 A
Selectable: 3 ,5 or 10 Wires
3 Lead wires ECG Cable
5 Lead wires ECG Cable
10 Lead wires ECG Cable
 5 mV
< 90 nA
4, 2, 1, 1/2, 1/4, Auto
1mV, 0.5 sec
"MONITOR"
0.5 - 24 Hz
"NORMAL"
0.5 - 40 Hz
"EXTENDED"
0.05-100 Hz
> 98 dB
< 30 µV RTI
> 5 MΩ
Duration
40 to 120 msec
Amplitude
0.25 to 5 mV
0.2 to 5 mV
15 - 300 BPM
for adult/Pediatric
15 - 350 BPM
for neonate
1% or 2 BPM
Reject up to 1.2 mV Amp.
Duration
0.1 - 2 msec
Amp
±2 to ± 700 mV (Without over/undershoot)
22-2
User Manual
18.5"
for Adult/Pediatric
for Neonate