Chapter 3: Alarm
User Manual
button, the alarm suspension status will be ended and the normal alarm status resumed
immediately.

3.6 Parameter Alarm

Alarm setting of each parameter can be found in its specific window. You can observe and set
the alarm limits and alarm features of each parameter in its specific window.
When a parameter alarm is 'OFF', this symbol "
" is displayed beside the parameter.
When parameter alarm is 'ON', alarm limits are displayed beside the parameter. If parameter
value exceeds the adjusted alarm limits, the alarm will be triggered and the following actions will
take place:
1. Alarm message is displayed on the screen.
2. The monitor beeps in its corresponding alarm level and volume.
3. Alarm indicator flashes.
NOTE:
For more specific information on alarms of each parameter, refer to Appendix II.
Warning
Alarm settings including priority, limits and volume should be done with regard to the
patient and environment conditions in a way that the patient life is not threatened and
reoccurrence of alarms is prevented.

3.7 When an alarm occurs

1. Check the patient's condition.
2. Recognize related alarms to each module.
3. Identify the alarm cause.
4. Press Silence button, if necessary.
5. After removing the alarm cause, enable the alarm sound.
You will find the alarm messages of each module in its own chapter.
3-6