Chapter 5: ECG Monitoring
NOTE:
-
You can choose aVR, aVL, aVF and V just when ECG is in 5-WIRE mode.
-
The leads V2, V3, V4, V5 and V6 can be observed only in 12-lead ECG mode.
NOTE:
Main lead is selectable in ECG menu. In pages that more than one trace of ECG
signal is displayed, the first trace is related to the main lead.
NOTE:
ST, ARR, Pace and HR are calculated from main lead that is displayed on the first
trace and can be adjusted in ECG menu.
NOTE:
Due to higher amplitude of signal voltage in leads II and V, it is recommended to
select one of these leads as main lead.
NOTE:
If an ECG waveform is not accurate while the electrodes are properly attached, try
to change the lead.