Chapter 9. Technical Specifications
Chapter 9. Technical Specifications
CLASSIFICATION
Protection against electroshock
Mode of operation
Harmful Liquid Proof Degree
Method of disinfection
Safety of anesthetic mixture
DISPLAY
Display
Waveforms
Numeric Parameters
Operation Method
Displayed data
ECG
Input Channel
Standard leads acquired
Sensitivity Selection
Filters
Calibration
Dynamic Range
Leakage Current
CMRR
Time Constant
Frequency Response
Pace
Standards
ECG Storage
Internal Memory
Recorder
Model
Print Method
Dots per line
Class I, Type CF Defibrillation proof (based on IEC 60601-1)
Continues operation equipment
Ordinary equipment, (without Liquid Proof)
Refer to chapter 10 for detail
Not suitable for use in the presence of a flammable anaesthetic
mixture with air or with oxygen or nitrous oxide.
TFT COLOR 800×480, 5"
12 Lead ECG/ Rhythm Lead
HR
Membrane Keys and Touch
Waveforms, Patient Information (Name and ID), Recording Speed,
Operation Mode, Filter, HR Value, Message
Simultaneous acquisition of all 12 leads/ Rhythm Lead
I, II, III, aVR, aVF, aVL, V1, V2, V3, V4, V5, V6
2.5, 5, 10, 20 mm/mV
Drift: on or off
HUM: 50 or 60 HZ, off
Low pass: 25, 35, 75, 150 HZ
EMG: on or off
1 mV
±5 mV
< 10 μA
> 98 dB
3.2 sec.
0.05~150 Hz
Detection & Rejection: 0.1~2 ms, ±2~±250 mV
Indication : 0.5~2 ms, ±2~±250 mV
IEC 60601-2-25:2011
Up to 100 Records
Thermal Printer
Thermal dot line printing
448 dots
44

Advertisement

Table of Contents
loading

Table of Contents