Chapter 10: NIBP Monitoring
Measuring pressure is more than
NIBP RANGE
upper limit (255mmHg)for adult
EXCEEDED
or (135mmHg) for neonate
NIBP
Arm movement, noisy signal or
EXCESSIVE
irregular pulse(e.g. arrhythmia)
MOTION
Measured pressure exceeds safe
software limit, 290 mmHg for
adult, 240 mmHg for pediatric
NIBP OVER
and 145mmHg for neonate.
PRESSURE
(NIBP SAADAT: measured
SENSED
pressure exceeds safe software
limit, 290 mmHg for adult and
150 mmHg for neonate)
Large motion artifact that
NIBP SIGNAL
saturates the amplifier's
SATURATED
amplitude handling capability.
NIBP
PNEUMATIC
Leakage during leak test
LEAK
Measurement time exceeds 3
NIBP TIME
minutes (2 minutes in CAS
OUT
module) for adults and pediatrics
or 90 seconds for neonates.
10-14
User Manual
Alarm level is set in NIBP
Window. By pressing ALARM
SILENCE, the message
background becomes gray and the
alarm is disabled and ignores this
fault.
Alarm level is set in NIBP
Window. By pressing ALARM
SILENCE, the message
background becomes gray and the
alarm is disabled and ignores this
fault.
Alarm level is set in NIBP
Window. By pressing ALARM
SILENCE, the message
background becomes gray and the
alarm is disabled and ignores this
fault.
Alarm level is set in NIBP
Window. By pressing ALARM
SILENCE, the message
background becomes gray and the
alarm is disabled and ignores this
fault.
Alarm level is set in NIBP
Window. By pressing ALARM
SILENCE, the message
background becomes gray and the
alarm is disabled and ignores this
fault.
Alarm level is set in NIBP Window.
By pressing ALARM SILENCE, the
message background becomes gray
and the alarm is disabled and ignores
this fault.