Chapter 24: Care and Cleaning
24.1 System Check
Before using the monitor,
■ Check if there is any mechanical damage in the system and accessories.
■ Check if all the power cable and accessories are firmly connected.
■ Check all the functions of keyboard and modules to make sure that the monitor is in proper
condition.
If you find any damage on the monitor, stop using the monitor on patient, and contact the
biomedical engineer of the hospital or local After Sale Service.
The overall check of the monitor, including the safety check, should be performed only by
qualified personnel.
All checks which need the monitor to be opened and safety and maintenance checks should be
performed by After Sales Service.
Note:
It is recommended that the system is calibrated by manufacturer every year, but it
has to be calibrated once every 2 years. In addition, the system lifetime is 10 years.
The medical canter can request the system calibration whenever the system accuracy
is in doubt.
Warning
If user does not follow a satisfactory maintenance schedule, the monitor may become
invalid, and human health may be endangered.
Note:
To ensure maximum battery life, it is recommended that, at least once a month, the
monitor runs on battery until it turns itself off and then recharged.